Operating Range:
0°C to 65°C standard
Ambient Temperature:
10°C to 40°C non-condensing
Stability / Repeatability:
± 0.1°C at constant load (even near ambient)
Cooling Capacity*:
275 Watts @ 20°C (20°C ambient)
Noise (at 1 meter):
< 63 dBA
Process Fluid:
PAO (Poly-Alpha-Olefin), Options available for Fluorinert / Galden or HFE
Process Fluid Fittings:
1/4" CP metal (CPM) or swage fittings
Pumps:
1, 2 or 3 lpm at 30 psig magnetically coupled gear pump
Tank Volume:
900 ml stainless steel tank with level sensor
Wetted Materials:
Aluminum, stainless steel and polymers
Fluid Filter:
5 μm external filter with grounding strap
Size (L x W x H):
13" x 11" x 13" (32 cm x 28 cm x 32 cm) without filter
Power Input:
Universal: 115-230 VAC, 50/60 Hz, 7-5 amp max
Communications:
Keypad or RS232 Interface
Alarms:
Temperature, fluid level, system or component
failure (display, RS232 and Dry Contact)
Standards:
TUV listed to UL, CAN/CSA and EN 61010-1,
CE 61010-1
* Note: Cooling capacity will vary with configuration